Peterson ME, Rishniw M, Bilbrough GE, et al.
J Feline Med Surg 2017:1098612×17708918.
Objectives The Catalyst One Chemistry Analyzer (IDEXX Laboratories) is a point-of-care instrument that can measure total thyroxine (TT4) by immunoassay. The aims of this study were to evaluate the analytic performance of the Catalyst TT4 assay in feline sera and to examine agreement of the Catalyst TT4 results with those measured by immunoassay at a veterinary reference laboratory. Methods Assay precision, reproducibility and linearity were evaluated for the Catalyst TT4 assay. For method comparison, TT4 concentrations in serum samples from 157 cats (127 hyperthyroid, 30 radioiodine-treated cats) were analyzed by both in-clinic and reference laboratory methods. Results The Catalyst TT4 demonstrated good precision and reproducibility (coefficients of variation 8.5%) and excellent linearity in the diagnostic range of 6-150 nmol/l. Differences between the two TT4 methods showed no proportional or fixed bias (Bland-Altman plots) but did demonstrate greater spread of values at higher TT4 concentrations. Statistical analysis of percent differences between methods indicated 95% limits of agreement of +/- 30%. When serum TT4 concentrations were classified as low, high or within the reference interval (12-50 nmol/l) for each assay, there was strong agreement (96.8%) in classification between methods. Conclusions and relevance The Catalyst TT4 assay provided precise serum TT4 concentrations in the 157 samples analyzed, which agreed well with results provided by a reference laboratory. Cats with Catalyst TT4 concentrations near decision thresholds (eg, normal vs high) should either have TT4 concentration repeated a few weeks later and/or undergo further testing (eg, free T4, serum thyroid-stimulating hormone, thyroid scintigraphy) to determine thyroid status.